Typical HVL measurements for different combinations of tube voltage and anode/filter combination / s-factors for clinically used spectra (Dance et al., 2000,).

HVL (mm Al) for anode/filter combination
kV Mo Mo Mo Rh Rh Rh W Rh W Ag W Al (0.5mm) W Al (0.7mm)
25 0.32 ±.02 0.38 ±.02 0.37 ±.02 0.50 ±.03 0.51 ±.03 0.34 ±.03 0.42 ±.03
28 0.35 ±.02 0.42 ±.02 0.42 ±.02 0.53 ±.03 0.58 ±.03 0.39 ±.03 0.49 ±.03
31 0.38 ±.02 0.45 ±.02 0.45 ±.02 0.56 ±.03 0.61 ±.03 0.44 ±.03 0.55 ±.03
34 0.40 ±.02 0.47 ±.02 0.47 ±.02 0.59 ±.03 0.64 ±.03 0.49 ±.03 0.61 ±.03
37       0.62 ±.03 0.67 ±.03 0.53 ±.03 0.66 ±.03
Target material Filter material Filter thickness (μm) S-Factor
Mo Mo 30 1.000
Mo Rh 25 1.017
Rh Rh 25 1.061
W Rh 50-60 1.042
W Ag 50-75 1.042