1 Introduction

Radiation dose optimization in medical imaging has long been a focal point in radiological practices, given the inherent trade-off between diagnostic accuracy and patient safety (Tsapaki, 2020). Computed tomography (CT) is a widely used diagnostic tool in radiology, providing high-resolution images of various anatomical structures and pathological conditions. However, CT exposes patients to relatively high doses of ionizing radiation compared to other radiological procedures. The rapid growth in CT utilization over the past few decades has raised concerns regarding the potential cancer risks associated with the cumulative radiation exposure in the population (Shubayr et al., 2023). Several studies have demonstrated a clear association between exposure to ionizing radiation from CT scans and increased lifetime attributable risk of cancer (Alashban and Shubayr, 2023). Therefore, radiation protection in CT is essential to ensure that the benefits of the examination outweigh the potential harms, and that the radiation dose is kept as low as reasonably achievable (ALARA) without compromising the diagnostic quality.

One of the practical tools for radiation protection in CT is the use of diagnostic reference levels (DRLs), which provide guidance on radiation dose optimization in medical imaging procedures (Rao et al., 2023; Jasieniak et al., 2023). DRLs are indicative values of radiation dose for a given procedure based on audit data from a large number of patients at a local, regional or national level (Alashban and Shubayr, 2022). DRLs represent the third quartile of the observed dose distribution for a given procedure based on audit data from numerous patients at a local, regional, or national level. DRLs are not dose limits, but rather benchmarks for comparison and optimization of patient exposure (Damilakis et al., 2023). The DRLs can be expressed in terms of dose indices such as CT dose index (CTDI) and dose length product (DLP), which reflect the output of the CT scanner (Kanal et al., 2017). DRLs can help to identify unusually high or low doses for a particular procedure, and to initiate actions to optimize the exposure parameters and the quality assurance program.

The DRL concept allows comparison of an individual facility’s typical radiation dose to the national benchmark for that examination. Regular nationwide surveys are required to update DRLs periodically to reflect evolving CT technology and practice (Nam et al., 2022; Abuzaid et al., 2020b). Many countries have implemented national DRLs (NDRLs) as a radiation protection tool to guide CT practice towards optimization (Kumsa et al., 2023; Rao et al., 2023). Several studies have evaluated the implementation of NDRLs in different countries and regions and reported significant dose reduction and optimization for various CT examinations (Korir et al., 2016; Skovorodko et al., 2022; Demb et al., 2017; Wachabauer et al. 2020).

The Saudi Food and Drug Authority (SFDA) recently established national DRLs (NDRLs) for CT examinations commonly performed in Saudi Arabia. However, the impact of implementing DRLs on patient radiation exposure has not been extensively evaluated in Saudi Arabia. Therefore, the aim of this study is to investigate the impact of the implementation of NDRLs on patient radiation dose  in terms of effective dose (ED)&9472; during chest CT examinations for adult patients in Saudi Arabia. Chest CT is one of the most frequently performed CT examinations, and it involves relatively high radiation doses, which make it a suitable candidate for dose optimization and audit.

2 Materials and methods

2.1 Study setting and data collection

Retrospectively, data were collected from four CT scanners (SOMATOM Definition AS 64-slice, SOMATOM Definition flash dual-source 128-slice, Siemens, Munich, Germany) at the Radiology Department in a major hospital in Jeddah, Saudi Arabia. The dataset comprises two-time frames: from March 2021 to February 2022 when NDRLs was not implemented yet; and from March 2022 to February 2023 when NDRLs has been incorporated into clinical settings. Prior to commencing of the study, medical ethics approval was obtained from the Research Ethics Committee (REC) of King Fahd Armed Forces Hospital (KFAFH) (REC 621). The collected data were anonymized, analyzed, and reported solely in an aggregate form.

2.2 Data extraction and parameters

The data was extracted from the Picture Archiving Communication System (PACS). The imaging parameters and dose metrics of CT chest imaging from adult patients were collected. The CT examinations included both contrasted and non-contrasted studies. Data comprised dose reports pertinent to the respective chest CT scan, including the scan protocol (i.e., kVp, mAs, scan length, DLP and CTDIvol).

2.3 Calculation method for effective dose

To estimate the ED from Chest CT scans, we utilized the DLP, and a region-specific conversion factor known as the k-factor. The ED calculation is based on the formula:

ED(mSv)=DLP(mGy·cm)×kfactor(mSvmGy·cm)

For our study, the k-factor for chest CT was sourced from the recently published data from a large registry/phantom library, taking into account scanner manufacturers (Chu et al, 2023), which is different from previous k-factors that were derived from theoretical models rather than real patient data. As stated by Chu et al., “the new coefficients offer reasonably reliable values for estimating effective dose.” The k-factor for chest CT for Siemens scanner is 0.040 mSv/(mGy · cm) (Chu et al., 2023). It should be noted that value of 0.040 is about 1.8 and 2.6 times the previous values of 0.020, 0.014, and 0.0146 from previous published k-factors (Huda et al., 2008; Deak et al., 2010; Shrimpton et al., 2006). In addition, the ED is not intended for dose to an individual; intended for populations.

2.4 Reference levels

Based on the NDRLs for chest CT, the reference levels for CTDIvol and DLP are set at 12 (mGy) and 430 (mGy · cm), respectively. These levels act as benchmarks to assess pass/fail outcomes in imaging procedures. For this study, a procedure recording a CTDIvol over 12 or a DLP over 430 would be classified as a failure. Conversely, if both CTDIvol and DLP values are at or below these thresholds, the procedure is deemed a pass. Applying these standards to the provided data, any case with a CTDIvol exceeding 12 is labeled as “Fail CTDIvol,” while those surpassing a DLP of 430 are marked as “Fail DLP.” Cases exceeding both limits are categorized as “Fail CTDIvol and DLP,” and those within the thresholds for both measures are classified as “Pass.”

2.5 Statistical analysis

Statistical analyses were performed using SPSS Statistics software (SPSS version 26, IBM, USA). Descriptive and inferential statistics were conducted. Percentage change was conducted for prior to and after the implementation of NDRLs for radiation parameters. Mann–Whitney U test was used to compare the ED prior to and after the implementation of NDRLs. A p value of 0.05 was set for significance.

3 Results

In our study, we analyzed data from a total of 1,991 chest CT scans. 429 scans (21.5%) were performed with contrast, while 1,562 scans (78.5%) were conducted without contrast. For collected data before the implementation of NDRLs, 221 of these scans (11.1%) were with contrast and 731 scans (36.7%) were without contrast. After NDRLs implementation, the number of scans with contrast was 208 (10.4%), while those without contrast were 831 (41.7%).

3.1 Results prior to NDRLs implementation

Prior to the implementation of NDRLs, the study documented varied radiation dose metrics for adult patients undergoing CT chest scans, with and without contrast, as shown in Table 1. In adults undergoing non-contrast CT, kVp was 118.27 ± 5.26, mAs was 1469 ± 578.84, CTDIvol was 8.82 ± 3.38 mGy, DLP was 309.33 ± 131.18 mGy.cm. For contrast scans, kVp, mAs, CTDIvol and DLP were 101.35 ± 7.86, 1604 ± 662.84, 6.33 ± 2.69 mGy, and 239.98 ± 115.19, respectively.

Table 1

Radiation dose parameters in adult patients undergoing CT chest scans prior to NDRLs implementation.

3.2 Results post-NDRLs implementation

After the implementation of NDRLs, the dose metrics for CT chest examinations showed changes across adult patients, and for both without and with contrast-enhanced scans, as shown in Table 2. For patients undergoing CT chest examinations without contrast, the kVp was 118.33 ± 5.52, the mAs was 1433.46 ± 530, the CTDIvol was 8.72 ± 2.33 mGy, and the DLP was 301.86 ± 124.32 mGy cm. For patients undergoing CT chest examinations with contrast-enhanced scans, the kVp was 100.76 ± 6.47, the mAs was 1614 ± 684, the CTDIvol was 6.20 ± 2.63 mGy, and the DLP was 233.8 ± 110.9 mGy cm.

Table 2

Radiation dose parameters in adult patients undergoing CT chest scans post-NDRLs implementation.

3.3 Comparison of radiation dose parameters

The implementation of NDRLs has resulted in various changes in radiation dose parameters across different patient categories, as illustrated in Table 3. In the cohort without contrast, the changes were slight, with a 0.05% increase in kVp, decreases in mAs by 2.42%, CTDIvol by 1.13%, DLP by 2.41%, and ED by 2.29%. Chest CT with contrast-enhanced scans resulted in a 0.58% decrease in kVp, a 0.62% increase in mAs, and decreases in CTDIvol by 2.05%, and DLP by 2.58%.

Table 3

Percentage change in radiation dose parameters for patients undergoing CT scans post-NDRLs implementation.

3.4 Pass/fail distribution post-NDRLs implementation

In Table 4, evaluating contrast-related procedures against NDRLs, 91.8% (191/208) passed with contrast and 81.5% (677/831) without contrast, yielding an overall pass rate of 83.5% (868/1039). CTDIvol failures were 2.9% (6/208) with contrast and 4.9% (41/831) without contrast; DLP failures were 1.9% (4/208) with contrast and 4.4% (20/831) without contrast. Also, 3.4% (7/208) of procedures with contrast and 11.2% (93/831) without contrast failed both CTDIvol and DLP, resulting in a combined failure rate of 9.6% (100/1039).

Table 4

Pass/fail distribution of contrast-related procedures based on NDRLs.

3.5 Changes in patient effective dose

Figure 1 represents the results of the ED before and after implementing NDRLs. Prior to introducing NDRLs, the ED for non-contrast CT scans was 12.37 ± 5.25 mSv, and for contrast scans was 9.6 ± 4.61 mSv. After the implementation of NDRLs, there were changes in the ED: Among adults, the ED for non-contrast CT scans showed a small decline of 2.43% to 12.07 ± 4.99 mSv, and for contrast scans, a more pronounced reduction of 6.77% to 8.95 ± 4.44 mSv. There were no statistically significant different (p > 0.05).

Comparing the distribution of ED before and after the implementation of NDRLs is presented in Figure 2. The ED distribution before the implementation of NDRLs appears roughly normally distributed, with a clear peak and a tail extending to the right, indicating that there is a wide range of higher EDs. The ED distribution after NDRLs implementation also appears normally distributed but shows a notable shift towards the left compared to the ‘before NDRLs’ distribution, with the peak at a lower dose, and the distribution is tighter, with fewer instances of higher EDs.

thumbnail Fig. 1

Effective doses by contrast use prior to and post-NDRLs implementation.

thumbnail Fig. 2

Distribution of Effective doses prior to and post-NDRLs implementation.

4 Discussion

The implementation of NDRLs is pivotal in standardizing radiation doses across medical imaging procedures, thereby enhancing patient safety by minimizing unnecessary exposure. NDRLs serve as a benchmark for radiological practices, promoting the optimization of dose protocols to align with best practice standards and ultimately improve patient care outcomes. In this study, implementing NDRLs has led to changes in radiation dose metrics for adult patients undergoing CT chest scans, with and without contrast.

The observed changes in radiation dose parameters for CT chest scans following the implementation of NDRLs suggest a potential impact on patient radiation exposure. The general decrease in mAs, CTDIvol, and DLP for non-contrast scans, and in kVp, CTDIvol, and DLP for contrast-enhanced scans, indicates a reduction in radiation dose. This aligns with previous research that has highlighted the role of NDRLs in optimizing patient dose levels (Abuzaid et al., 2020a). However, the slight increase in mAs for contrast-enhanced scans and in kVp for non-contrast scans warrants further investigation. The negligible increase, although counterintuitive, may be indicative of adjustments made to maintain image quality. It is crucial to balance the need for diagnostic image quality with the goal of minimizing radiation exposure. Studies have reported a reduction in radiation exposure without compromising image quality (Kumamaru et al., 2016; Aberle et al., 2020; Sarma et al., 2022). These findings highlight the importance of continuous monitoring and adjustment of CT scan parameters in line with NDRLs to ensure optimal patient safety.

The findings presented in Table 4 offer significant insights into the pass/fail distribution of contrast-related procedures against NDRLs. The pass/fail distribution based on NDRLs shows that most cases (83.5%) passed the criteria. This is consistent with a range of studies that have found that most cases pass the NDRLs, suggesting that radiation doses in medical imaging are generally below the recommended NDRLs (Wachabauer et al, 2020; Bijwaard et al., 2017).

The higher pass rate in procedures with contrast (91.8%) compared to those without (81.5%) suggests that the use of contrast in diagnostic imaging might be associated with better adherence to NDRL standards. This could be due to more precise imaging achieved with contrast agents, leading to optimized radiation doses. Subsequently, the ED for non-contrast CT scans showed a small decline of 2.43%, and for contrast scans, a more pronounced reduction of 6.77% (non-significant, p > 0.05). However, even small reductions in radiation exposure can be beneficial, especially when considering the cumulative effects of radiation from multiple CT scans over time. The use of NDRLs has been particularly effective in reversing the upward trend in radiation dose for chest CT scans (Rawat et al, 2015; Yang, 2020).

Overall, prior to NDRLs, the ED distribution shows a wider range and a right-skewed tail of higher EDs, while post-NDRLs, the distribution shifts leftward with a lower peak dose and a tighter distribution, indicating lower average EDs with reduced variability and fewer high-dose instances. Moreover, the findings highlight higher NDRL compliance in procedures with contrast and indicates areas for dose optimization improvement in procedures without contrast. Ensuring adherence to NDRLs in these procedures is essential for patient safety, the overall effectiveness of diagnostic imaging and regulatory compliance. This might involve revisiting protocol standards, enhancing training, or incorporating advanced imaging technologies that optimize dose without compromising diagnostic quality (Greenwood et al., 2015; Guerra et al. 2019; Tsapaki, 2020).

A critical aspect to consider when evaluating the impact of NDRLs on radiological practices is the temporal proximity between the NDRL implementation and the subsequent evaluation period. The interval observed in this study is notably brief, suggesting that the window for instituting substantial modifications in clinical practices was limited. This constraint highlights a potential impediment to realizing more significant reductions in patient exposure to radiation. Furthermore, the endeavor to decrease patient exposure through changes in radiological practices extends beyond mere procedural adjustments; it fundamentally necessitates awareness-raising actions and a commitment to continuous professional education (Bertho and Habib, 2023). The importance of ongoing education for healthcare professionals in facilitating the adoption of safer, more effective radiological techniques that align with NDRL guidelines, emphasizing that such educational initiatives are vital for sustaining long-term improvements in patient safety and care quality.

4.1 Limitations of the study

This study has some limitations that should be considered when interpreting the results. This study only focused on chest CT examinations, which may not reflect the radiation dose variation and optimization for other CT examinations in different body regions. Future studies should include more body organs and larger and more representative samples,

5 Conclusion

The adoption of the NDRLs has resulted in reduction in radiation doses during CT chest examinations. The ED declined by 2.29% for non-contrast and 6.63% for contrast CT scans A better adherence to NDRL standards with contrast use, as evidenced by a higher pass rate than non-contrast scans (91.8% vs. 81.5%). These findings suggest a regular monitoring and review for CT procedures to avoid unintended consequences. Therefore, this study recommends that CT users should follow the NDRLs as a guidance, but also consider other factors such as patient characteristics, protocol settings, and quality assurance program to ensure that the radiation dose is kept as low as reasonably achievable (ALARA) without compromising the diagnostic quality.

Acknowledgments

The authors would also like to extend their sincere appreciation to the Researcher supporting program at King Saud University, Riyadh, for funding this work under the project number (RSP2024 R328).

Funding

This study was supported by the Researcher Supporting Program at King Saud University, Riyadh, under project number RSP2024 R328.

Conflicts of Interest

The author declares that he has no conflicts of interest in relation to this article.

Data availability statement

The data that support the findings of this study are available from the corresponding author upon reasonable request.

Author contribution statement

R. Sindi: Conceptualization, Methodology, Investigation, Writing – Original Draft, Project Administration. B. Al-Shamrani: Data curation, Investigation. A. Bana: Data curation, Investigation. F. Al-Qurashi: Data curation, Investigation. M. Al-Qarhi: Data curation, Investigation. B. Al-Shehri: Resources, Investigation. R. Al-Otaibi: Investigation, Formal Analysis. S. Aldawood: Validation, Visualization, Funding Acquisition. N. Shubayr: Conceptualization, Formal Analysis, Writing – Review & Editing, Writing – Original Draft, Supervision.

Ethics approval

All ethical considerations were adhered to, including obtaining institutional review board approval. Confidentiality and anonymity were strictly maintained.

Informed consent

This study did not require informed consent.

References

  • Aberle C, Ryckx N, Treier R, Schindera S. 2020. Update of national diagnostic reference levels for adult CT in Switzerland and assessment of radiation dose reduction since 2010. Eur Radiol 30: 1690–1700. [CrossRef] [PubMed] [Google Scholar]
  • Abuzaid MM, Elshami W, El Serafi A, Hussien T, McConnell JR, Tekin HO. 2020a. Toward national Ct diagnostic reference levels in the United Arab Emirates: a multicenter review of Ct dose index and dose length product. Radiat Prot Dosimetry 190: 243–249. [CrossRef] [PubMed] [Google Scholar]
  • Abuzaid MM, Elshami W, Tekin HO, Ghonim H, Shawki M, Salama DH. 2020b. Computed tomography radiation doses for common computed tomography examinations: a nationwide dose survey in United Arab Emirates. Insights Imaging 11: 88. [CrossRef] [PubMed] [Google Scholar]
  • Alashban Y, Shubayr N. 2022. Establishing diagnostic reference levels for CT examinations in the south region of Saudi Arabia. Radiat Phys Chem 201: 110407. [CrossRef] [Google Scholar]
  • Alashban Y, Shubayr N. 2023. Probability of induced cancers related to computed tomography examinations. Radiat Phys Chem 210: 111020. [CrossRef] [Google Scholar]
  • Bertho J-M., Habib Geryes B. 2023. La radioprotection est une attitude. Radioprotection 58: 77–78. [CrossRef] [EDP Sciences] [Google Scholar]
  • Bijwaard H, de Vries G, Scheurleer J, Roding T, Erenstein H, Ravensbergen W, Haarmans-Jonkman S, van Welie F. 2017. Compliance to diagnostic reference levels for radiation exposure in common radiological procedures in Dutch hospitals: a nation-wide survey carried out by medical imaging students. Radiography (Lond) 23: 197–201. [CrossRef] [PubMed] [Google Scholar]
  • Chu PW, Kofler C, Haas B, Lee C, Wang Y, Chu CA,Stewart C, Mahendra M, Delman BN, Bolch WE, Smith-Bindman R. 2023. Dose length product to effective dose coefficients in adults. Eur Radiol [Google Scholar]
  • Damilakis J, Frija G, Brkljacic B, Vano E, Loose R, Paulo G, Brat H, Tsapaki V, European Society of R. 2023. How to establish and use local diagnostic reference levels: an ESR EuroSafe Imaging expert statement. Insights Imag 14: 27. [CrossRef] [Google Scholar]
  • Deak PD, Smal Y, Kalender WA. 2010. Multisection CT protocols: sex- and age-specific conversion factors used to determine effective dose from dose-length product. Radiology 257: 158–166. [CrossRef] [PubMed] [Google Scholar]
  • Demb J, Chu P, Nelson T, Hall D, Seibert A, Lamba R, Boone J, Krishnam M, Cagnon C, Bostani M, Gould R, Miglioretti D, Smith-Bindman R. 2017. Optimizing radiation doses for computed tomography across institutions: dose auditing and best practices. JAMA Intern Med 177: 810–817. [CrossRef] [PubMed] [Google Scholar]
  • Greenwood TJ, Lopez-Costa RI, Rhoades PD, Ramirez-Giraldo JC, Starr M, Street M, Duncan J, McKinstry RC. 2015. CT dose optimization in pediatric radiology: a multiyear effort to preserve the benefits of imaging while reducing the risks. Radiographics 35: 1539–1554. [CrossRef] [PubMed] [Google Scholar]
  • Guerra VH, Nersissian DY, Melo CS, Vasconcellos C,Freitas RG, Sawamura M, Gebrim E, Costa PR. 2019. Paediatric CT dose optimization in a general hospital. Rev Bras Física Méd 13: 138–144. [Google Scholar]
  • Huda W, Ogden KM, Khorasani MR. 2008. Converting dose-length product to effective dose at CT. Radiology 248:995–1003. [CrossRef] [PubMed] [Google Scholar]
  • Jasieniak J, Kuchcinska A, Podgorska J, Cieszanowski A. 2023. Summary of radiation dose management and optimization: comparison of radiation protection measures between Poland and other countries. Pol J Radiol 88: e12–e21. [CrossRef] [Google Scholar]
  • Kanal KM, Butler PF, Sengupta D, Bhargavan-Chatfield M, Coombs LP, Morin RL. 2017. U.S. diagnostic reference levels and achievable doses for 10 adult CT examinations. Radiology 284: 120–133. [CrossRef] [PubMed] [Google Scholar]
  • Korir GK, Wambani JS, Korir IK, Tries MA, Boen PK. 2016. National diagnostic reference level initiative for computed tomography examinations in Kenya. Radiat Prot Dosimetry 168: 242–252. [PubMed] [Google Scholar]
  • Kumamaru KK, Kogure Y, Suzuki M, Hori M, Nakanishi A, Kamagata K, Hagiwara A, Andica C, Ri K, Houshido N, Aoki S. 2016. A strategy to optimize radiation exposure for non-contrast head CT: comparison with the Japanese diagnostic reference levels. Jpn J Radiol 34: 451–457. [CrossRef] [PubMed] [Google Scholar]
  • Kumsa MJ, Nguse TM, Ambessa HB, Gele TT, Fantaye WG, Dellie ST. 2023. Establishment of local diagnostic reference levels for common adult CT examinations: a multicenter survey in Addis Ababa. BMC Med Imag 23: 6. [CrossRef] [Google Scholar]
  • Nam S, Park H, Kwon S, Cho PK, Yoon Y, Yoon SW, Kim J. 2022. Updated national diagnostic reference levels and achievable doses for CT protocols: A national survey of korean hospitals. Tomography 8: 2450–2459. [CrossRef] [PubMed] [Google Scholar]
  • Rao S, Sharan K, Sukumar S, Chandraguthi SG, Nisha Dsouza R, David LR, Ravichandran S, Uzun B, Kadavigere R, Uzun Ozsahin D. 2023. Systematic review on diagnostic reference levels for computed tomography examinations in radiation therapy planning. Diagnostics (Basel) 13. [PubMed] [Google Scholar]
  • Rawat U, Cohen S, Levsky JM, Haramati LB. 2015. ACR white paper-based comprehensive dose reduction initiative is associated with a reversal of the upward trend in radiation dose for chest CT. J Am Coll Radiol 12: 1251–1256. [CrossRef] [PubMed] [Google Scholar]
  • Sarma AD, Sharma J, Singha M. 2022. A review on diagnostic reference levels for adult patients undergoing chest (coronary angiography) computed tomography scan in North-East India. Asian Pacific J Health Sci 9: 2020. [Google Scholar]
  • Shrimpton PC, Hillier MC, Lewis MA, Dunn M. 2006. National survey of doses from CT in the UK: 2003. Br J Radiol 79: 968–980. [CrossRef] [PubMed] [Google Scholar]
  • Shubayr N, Alashban Y. 2023. Estimation of radiation doses and lifetime attributable risk of radiation-induced cancer in the uterus and prostate from abdomen pelvis CT examinations. Front Public Health 10. [CrossRef] [PubMed] [Google Scholar]
  • Skovorodko K, Komiagiene R, Maciusovic M, Gilys L, Vajauskas D, Grigoniene V, Ziliukas J, Raudoniene J, Griciene B. 2022. Nationwide survey on radiation doses received by patients in nuclear medicine imaging procedures. J Radiol Prot 42. [Google Scholar]
  • Tsapaki V. 2020 Radiation dose optimization in diagnostic and interventional radiology: current issues and future perspectives. Phys Med 79: 16–21. [CrossRef] [PubMed] [Google Scholar]
  • Wachabauer D, Rothlin F, Moshammer HM, Homolka P. 2020. Diagnostic reference levels for computed tomography in Austria: a 2018 nationwide survey on adult patients. Eur J Radiol 125: 108863. [CrossRef] [PubMed] [Google Scholar]
  • Yang CC. 2020. Evaluation of impact of factors affecting CT radiation dose for optimizing patient dose levels. Diagnostics (Basel) 10: 787. [CrossRef] [PubMed] [Google Scholar]

Cite this article as: Sindi R, Al-Shamrani B, Bana A, Al-Qurashi F, Al-Qarhi M, Al-Shehri B, Al-Otaibi R, Aldawood S, Shubayr N. 2024. Impact of implementing national diagnostic reference levels on radiation dose optimization in adult chest CT scans: a comparative analysis. Radioprotection 59(2): 111–116

All Tables

Table 1

Radiation dose parameters in adult patients undergoing CT chest scans prior to NDRLs implementation.

Table 2

Radiation dose parameters in adult patients undergoing CT chest scans post-NDRLs implementation.

Table 3

Percentage change in radiation dose parameters for patients undergoing CT scans post-NDRLs implementation.

Table 4

Pass/fail distribution of contrast-related procedures based on NDRLs.

All Figures

thumbnail Fig. 1

Effective doses by contrast use prior to and post-NDRLs implementation.

In the text
thumbnail Fig. 2

Distribution of Effective doses prior to and post-NDRLs implementation.