Open Access

Table 2

Combined standard uncertainties of technical parameters of neutron reference fields available in Vietnam.

Neutron reference field Quantity Combined standard uncertainty, k = 1 (%)
241Am-Be ϕ 2.1
  H˙*(10) 4.5
  E¯ 3.0
  h* 5.0

241Am-Be (15PE) Φ 3.2
241Am-Be (20PE) H˙*(10) 5.1
241Am-Be (25PE)
241Am-Be (30PE) E¯ 4.5
241Am-Be (35PE) h* 6.0
239Pu-Be    

Notation: Value of uncertainty budget, ub (%):
Bb(Eb): 1.5% (stated in a certificate);
l: 1.0% (maximum estimated);
FA(l): 1.0% (based on influenced factors in calculation dependent on different total cross section uncertainty);
F1(90): 0.3% (maximum investigated);
h*(Eb): 4.0% (stated in (ICRU, 1998));
Mi: 1.0% (targeted in measurements);
Rib(Eb): 3.0% (estimated).
For direct component of neutrons from a bare 241Am-Be source (ISO-8529 criteria):
Uncertainty budgets taken into account for standard uncertainty of ϕ: ϕb(Eb), l, FA(l), F1(90) ;
Uncertainty budgets taken into account for standard uncertainty of H˙*(10): ϕb, h*(Eb);
Uncertainty budgets taken into account for standard uncertainty of E¯: ϕb
Uncertainty budgets taken into account for standard uncertainty of h*: ϕ; H˙*(10)
For total component of neutrons from other sources (ISO-12789 criteria):
Uncertainty budgets taken into account for standard uncertainty of ϕ: Mi, Rib(Eb)
Uncertainty budgets taken into account for standard uncertainty of H˙*(10): ϕ, h*(Eb);
Uncertainty budgets taken into account for standard uncertainty of E¯:ϕ
Uncertainty budgets taken into account for standard uncertainty of h*: ϕ; H˙*(10)

Current usage metrics show cumulative count of Article Views (full-text article views including HTML views, PDF and ePub downloads, according to the available data) and Abstracts Views on Vision4Press platform.

Data correspond to usage on the plateform after 2015. The current usage metrics is available 48-96 hours after online publication and is updated daily on week days.