Accès gratuit
Table 1
Technical parameters for both systems.
Application | O-ring system | C-arm system |
---|---|---|
2D imaging | Yes | Yes |
3D imaging | Yes | Yes |
Slice width | 0.83 mm | 0.312 or 0.5 mm |
Standard 3D volume | Cylinder 20 cm (diameter)*15 cm (height) | 16*16*16 cm3 |
Compatible navigation | Medtronic | Brainlab – Striker |
Pulse frequency | 30 | 1, 2, 4, 8, 12.5, 25 |
Choice of pulse rate | Fixed to 30 pps | Variable |
Physical system | ||
Technology of detector | Amorphous Silicon | Amorphous Silicon |
Scintillator | Caesium Iodide | Caesium Iodide |
Detector pixel length | 194 μm | 194 μm |
3D spatial resolution | 512*512*192 voxels | 3203voxels = cuts to 0.32 mm 5123voxels = cuts to 0.5 mm |
Detector dimensions | 39.7*29.8 cm2 | 29.8*29.8 cm2 |
Resolution | 1.29 lp/mm | 2.6 lp/mm |
DQE (detective quantum efficiency) to 0 lp/mm | 77% | 70 ± 5% |
X-ray generator power | 32 kW | 25 kW |
kV Tube rating | 40–120 | 40–120 |
Regulation of mA | 10, 12.5, 16, 20, 25, 32, 40, 50, 64, 80, 100, 125, 160, 200, 250, 320 | 1.5 to 250 |
Distance focal – detector | 117 cm | 104.5 cm |
Opening | 96.5 cm | 84 cm |
Weight | 885 kg | 415 kg |
Angulation for 3D acquisition | 360° | 180° |
Total filtration | > 4 mm Al | ≥ 4.3 mm Al, including 0.1 mm Cu |
Dosimetric quantities provided | ||
CTDI | [mGy] | |
DLP | [mGy cm] | |
DAP | [Gy cm2] |