Parameter of neutron calibration field in Viet Nam following international standard
Institute for Nuclear Science and Technology, 179 Hoang Quoc Viet − Cau Giay, Hanoi, Viet Nam
* Corresponding author: LnThiem@vinatom.gov.vn
Received:
12
January
2024
Accepted:
27
May
2024
This paper presents the technical parameters of neutron calibration fields available in Viet Nam for calibrations of neutron measuring meters (i.e., neutron ambient dose equivalent/rate meters and/or neutron personal dose equivalent/rate meters − referred to neutron meters). The technical parameters of neutron calibration fields, in this work, consist of physical quantity (i.e., the integrated neutron fluence rate over a neutron spectrum, ϕ); operational dosimetric quantities (i.e., ambient dose equivalent rate, Ḣ*(10)) and radiation quantities (i.e., fluence-spectrum-averaged energies, Ē; fluence-to-ambient dose equivalent conversion coefficient, h*). The technical parameters of neutron reference fields were reported for different radiation beam qualities following the ISO-8529 and the ISO-12789 series (i.e., for direct neutron component from a neutron reference fields of a bare 241Am-Be, complied with the ISO-8529 series criteria; and for total neutron component from simulated workplace neutron reference fields of a 241Am-Be source moderated by different polyethylene spheres and from a neutron reference field of a bare 239Pu-Be source, complied with the ISO-12789 series criteria). The technical parameters of the ISO-8529 neutron reference field of the bare 241Am-Be source were theoretically calculated based on the total neutron source strength and presented for direct neutron component. In addition, the technical parameters of the ISO-12789 simulated workplace neutron reference fields of the 241Am-Be source moderated by different polyethylene spheres as well as of the 239Pu-Be source were characterized using the Bonner sphere spectrometer system accompanying with the FRUIT neutron spectrum unfolding code for the total neutron component. The technical parameters as well as their combined standard uncertainties of neutron fields were also presented in conjunction with different distances from the sources’ centers.
Key words: Quantities / operational dosimetric / reference fields
© N.T. Le, Published by EDP Sciences 2025
This is an Open Access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
Les statistiques affichées correspondent au cumul d'une part des vues des résumés de l'article et d'autre part des vues et téléchargements de l'article plein-texte (PDF, Full-HTML, ePub... selon les formats disponibles) sur la platefome Vision4Press.
Les statistiques sont disponibles avec un délai de 48 à 96 heures et sont mises à jour quotidiennement en semaine.
Le chargement des statistiques peut être long.