Calculation of self-absorption and coincidence summing correction factors for the extended sources using GEANT4
Department of Nuclear Science and Technology, School of Energy and Power Engineering, Xi’an Jiaotong University,
Xi’an
710049, PR China
* Corresponding author: hechaohui@xjtu.edu.cn
Received:
27
September
2018
Accepted:
12
February
2019
A detailed study of the full energy peak efficiency of a high purity germanium (HPGe) detector including the effect of source self-absorption and coincidence summing was performed using Monte Carlo simulation, as it is difficult and time-consuming to measure the full energy peak efficiency experimentally. Cylindrical water composition source was simulated with different characteristics, covering the energy range from 60 to 1836 keV. Self-absorption correction factors (SAFcal) were calculated for two source volumes and obtained good agreement with the experimental results except for (60Co and 88Y) nuclides. The simulation was performed for various samples with different densities and observed their effects on the full energy peak efficiency value of the detector. In the case of extended volumetric source, the coincidence summing correction factors (CSFcal) for two nuclides (60Co and 88Y) were estimated with the GEANT4 simulation toolkit. The effect of correction factors on different cylindrical source volumes was also investigated. With the self-absorption and coincidence summing effect, the best agreement was achieved between simulated and experimental results with discrepancy less than 2% for all of the radionuclides included in two source volumes.
Key words: GEANT4 / HPGe detector / full energy peak efficiency / self-absorption / coincidence summing / cylindrical sources
© EDP Sciences 2019
Les statistiques affichées correspondent au cumul d'une part des vues des résumés de l'article et d'autre part des vues et téléchargements de l'article plein-texte (PDF, Full-HTML, ePub... selon les formats disponibles) sur la platefome Vision4Press.
Les statistiques sont disponibles avec un délai de 48 à 96 heures et sont mises à jour quotidiennement en semaine.
Le chargement des statistiques peut être long.