Open Access

Table 6

Dosimetric indicators found for the four types of examination.

CT protocols Dosimetric indicators 75th percentile
this study
75th percentile
(Benamar et al., 2023)
75th percentile
(El Mansouri et al., 2022)
75th percentile
France
(2019–2021)
(IRSN 2023)
DRL Australia
(Lee et al., 2020)
DRL
Allemagne (Schegerer et al., 2019)
DRL Japan
(Abe et al., 2020)
DRL Nigeria (Ekpo et al., 2018)
Head CTDIvol (mGy) 56.7 57,7 57.4 39.6 52 60 77 61
DLP (mGy.cm) 1166.76 1250,4 1020 734 880 850 1350 1310
Chest CTDIvol(mGy) 11.82 11,1 12.3 6.6 10 10 13 17
DLP (mGy.cm) 450 392,2 632 248 390 350 510 735
Abdomen CTDIvol (mGy) 16.1 11,3 10.9 9.5 13 15 18 20
DLP (mGy.cm) 866.69 517,1 714 475 600 700 880 1486
CAP CTDIvol (mGy) 12.32 11,6 9.1 11 13 16
DLP (mGy.cm) 919.8 833,27 620 940 1000 1200