Measured mean breast surface dose during thoracic CT scans with and without the constructed shields.

Shield composition ESD (mGy) Dose reduction (%)
No shielding 15.03 ± 2.13
50% Cu–50%BaSO4 7.93 ± 0.63 47.23
10% Cu–90% BaSO4 8.03 ± 1.52 46.57
100% BaSO4 13.64 ± 1.35 14.17
100% Bi 7.26 ± 0.71 51.69

Dose reduction (%) = 100 × [(Measured dose without shield − Measured dose in presents of shield)/Measured dose without shield].