Table 1
Clinical protocol parameters used in routine imaging for each of the three modalities.
CT | CBCT | OPT | ||
---|---|---|---|---|
Facial bones | Dental | “Full” – coverage facial bones | “Zoom” – dental | Standard |
120 kVp | 110 kVp | 66 kVp | ||
240 mAs | 90 mAs | [1–20] mA/3.6 s | 9 mA/16 s | |
0.55 pitch | 0.8 pitch | |||
256 mm | 180 mm | (D*H) 15*15 cm2 | (D*H) 12*8 cm2 | (D*H) 30*14 cm2 |
26.9 mGy | 12.9 mGy | 3.0 mGy | 2.5 mGy | 103.5 mGy.cm2 |
The values indicated in: first row correspond to kilovolt; second row correspond to milliAmpere second; third row correspond to pitch (only for CT); fourth row correspond to Field Of View (“D” stands for diameter, “H” stands for height.); last row correspond to the volume computed tomography dose index CTDIvol for the CT and CBCT modalities and to the dose area product (DAP) for the OPT modality, respectively.
Current usage metrics show cumulative count of Article Views (full-text article views including HTML views, PDF and ePub downloads, according to the available data) and Abstracts Views on Vision4Press platform.
Data correspond to usage on the plateform after 2015. The current usage metrics is available 48-96 hours after online publication and is updated daily on week days.