Open Access

Table 2

Participants response regarding different aspects.

Questions Patient parameters False True
Routine CT scanning parameters (kVp, mAs, slice thickness, pitch, reconstruction algorithms, etc.) should be changed according to which of the following: Patient size 4 (1.2%) 324 (98.8%)
Anatomical region 36 (11%) 292 (89%)
Study indication 68 (20.7%) 260 (79.3%)
Patient age 56 (17.1%) 272 (82.9%)
Regarding Automated Tube Current Modulation (ATCM) ATCM has been shown to decrease patient dose on average 32 (9.8%) 296 (90.2%)
ATCM can increase the patient dose in the pelvic region 116 (35.4%) 212 (64.6%)
ATCM should not be used in the presence of metallic implants 184 (56.1%) 144 (43.9%)
ATCM is affected by centering of the patient within the gantry 52 (15.9%) 276 (84.1%)
Regarding the noise setting (Noise index: GE/ standard deviation: Toshiba/ Effective mAs: Siemens+ Philips) The non-contrast phase of an abdomen scan required the same noise setting (i.e., mAs setting) as the contrast phase 228 (69.5%) 100 (30.5%)
Readers can tolerate less noise with obese patients 172 (52.4%) 156 (47.6%)
Readers can tolerate more noise with paediatrics 108 (32.9% 220 (67.1%)
The noise index should be changed when changes in patient size are extreme 24 (7.3%) 304 (92.7%)
Reducing the kVp from 120 to 100 kVp for angiographic CT procedures (all other parameters being kept constant Reduces the radiation dose 28 (8.5%) 300 (91.5%)
Reduces the image contrast 124 (37.8%) 204 (62.2%)
Increases the image noise 104 (31.7%) 224 (68.3%)
Increases the vessel enhancement 112 (34.1%) 216 (65.9%)
Regarding the ’Pitch’ (table movement per rotation/ total nominal beam width) Pitch may impact on image quality and patient dose 20 (6.1%) 308 (93.9%)
Higher table speeds result in an increase in slice sensitivity profile and thus effective slice thickness, reducing the z-axis resolution 72 (22%) 256 (78%)
Spiral artefacts are reduced at lower pitch settings 80 (24.4%) 248 (75.6%)
For single helical CT, the higher the pitch, the lower the dose 68 (20.7%) 260 (79.3%)
Regarding slice thickness (selected beam width/ collimation) Increasing the slice thickness increases the spatial resolution 184 (56.1%) 144 (43.9%)
Increasing the slice thickness decreases the dose 72 (22.0%) 256 (78..0%)
Decreasing the slice thickness reduces "partial volume" artefacts 68 (20.7%) 260 (79.3%)
Decreasing the slice thickness will increase the scan time 176 (53.7%) 152 (46.3%)

Les statistiques affichées correspondent au cumul d'une part des vues des résumés de l'article et d'autre part des vues et téléchargements de l'article plein-texte (PDF, Full-HTML, ePub... selon les formats disponibles) sur la platefome Vision4Press.

Les statistiques sont disponibles avec un délai de 48 à 96 heures et sont mises à jour quotidiennement en semaine.

Le chargement des statistiques peut être long.