Advancing patient safety in interventional radiology: effective dose evaluation and LDRL
1
Medical Physics and Radioprotection Nucleus, Botucatu Medical School, Clinics Hospital, Botucatu, Brazil
2
Institute of Bioscience, São Paulo State University, Botucatu, Brazil
3
Medical Physics of Radiodiagnosis, Hospital de Clínicas, Botucatu Medical School, Botucatu, Brazil
4
Department of Infectious Diseases, Dermatology, Diagnostic Imaging and Radiotherapy/Botucatu Medical School, São Paulo State University, Botucatu, Brazil
* Corresponding author: diana.pina@unesp.br
Received:
8
May
2024
Accepted:
31
July
2024
This study highlights the significance of radioprotection in interventional radiology, focusing on the assessment of Effective Dose values to which patients are exposed. To facilitate this, we developed a semi-automatic software employing Optical Character Recognition technology. This software analysed 1334 interventional procedures conducted between January 2021 and September 2023 at a large Brazilian hospital. The database includes the type of procedure, Air Kerma value, Kerma-Area Product, number of cineangiography images, and exposure duration, all realized on an Artis Zee® (Siemens Healthineers, Germany) C-arm. We established Local Diagnostic Reference Levels (LDRLs) and calculated Effective Dose values using the PCXMC 2.0 Monte Carlo simulation software. These calculations were based on data from the hospital and LDRLs determined in subsequent studies. The findings reveal that both the LDRLs and the calculated Effective Dose generally fall below the values reported in existing literature. This underscores the efficacy of our methodology in enhancing radioprotection and dose monitoring within the studied hemodynamics sector.
Key words: radiation protection / fluoroscopy / radiation dose
© M. Alvarez et al., Published by EDP Sciences 2025
This is an Open Access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
Les statistiques affichées correspondent au cumul d'une part des vues des résumés de l'article et d'autre part des vues et téléchargements de l'article plein-texte (PDF, Full-HTML, ePub... selon les formats disponibles) sur la platefome Vision4Press.
Les statistiques sont disponibles avec un délai de 48 à 96 heures et sont mises à jour quotidiennement en semaine.
Le chargement des statistiques peut être long.