Article
Benefits and challenges of standard ceiling-mounted surface guided radiotherapy systems for breast treatments on Halcyon™
1
ORLAM Group, Department of Radiation Oncology,
44 rue Ambroise Paré,
71000
Mâcon, France
2
Vision RT Ltd., Dove House,
Arcadia Avenue,
London
N3 2JU, United Kingdom
* Corresponding author: Nguyen.daniel@yahoo.fr
Received:
26
March
2021
Accepted:
6
September
2021
The present work investigates the benefits and challenges of applying surface guided radiation therapy (SGRT) to breast cancer treatments on the Halcyon™ (Varian, USA). Inter-fraction setup accuracy and treatment time were assessed by comparing CBCT shifts registered following SGRT-based vs. standard tattoo-based patient setup in two randomly selected groups of ten breast patients. Next, using a torso phantom and a volunteer, intra-fraction motion tracking accuracy and surface coverage were assessed for 3 adjusted central ceiling camera positions. SGRT was shown to significantly reduce setup errors (mostly within ± 2 mm) compared to tattoos (up to 2.1 cm) in all translational directions (p-value < 0.001). Treatment throughput and efficiency were also significantly improved with SGRT (p = 0.038). Meanwhile, intra-bore surface coverage with an adjusted central SGRT camera proved insufficient, suffering from patient self-occlusions (invisible body parts occluded by patients’ own morphology such as breasts, bellies, arms, etc.) and bore-induced camera obstructions. Tracking accuracy remained satisfactory (sub-0.5 mm) but 6 degrees-of-freedom motion monitoring, critical in stereotactic radiosurgery, stereotactic body radiation therapy and deep inspiration breath hold techniques and clinical applications, was not possible. Standard ceiling-mounted SGRT systems reduce inter-fraction breast setup errors and treatment duration while intra-fraction motion tracking is insufficient for O-ring linear accelerators.
Key words: breast cancer / Halcyon™ / surface guidance / patient positioning / treatment time / intra-bore tracking
© The Authors, published by EDP Sciences 2021
This is an Open Access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
Les statistiques affichées correspondent au cumul d'une part des vues des résumés de l'article et d'autre part des vues et téléchargements de l'article plein-texte (PDF, Full-HTML, ePub... selon les formats disponibles) sur la platefome Vision4Press.
Les statistiques sont disponibles avec un délai de 48 à 96 heures et sont mises à jour quotidiennement en semaine.
Le chargement des statistiques peut être long.