Effect of electric and magnetic fields on operation of insulin pumps under 400 kV power lines
1
Clinical Physiology and Neurophysiology Unit, The North Karelia Central Hospital,
Tikkamäentie 16,
80210
Joensuu, Finland
2
Tmi Rauno Pääkkönen,
Tampere, Finland
3
Fingrid Oyj,
Helsinki, Finland
* Corresponding author: leenakorpinen@gmail.com
Received:
1
October
2019
Accepted:
3
November
2019
The aim was to study the operation of insulin pumps under a 400 kV transmission line (two test places) and possible disturbances that the lines could have caused. Three different insulin pumps were attached one at a time to the subject’s clothes. The diabetes nurse started the pumps before the tests to ensure the correct settings were being utilized. After walking under the power lines when meters were running, she tested if the meters had operated properly. At the first test site (A), the electric field was 5.0–5.1 kV/m and the magnetic field 8.6–10.9 μT, and in the second test place (B), the fields were 7.7–8.5 kV/m and 5.7–9.2 μT. The pumps worked perfectly: no disruption was detected in the flow, display or menu movement, and the events were registered correctly. Only the remote control, which worked well before and after the test, could not reliably receive additional doses. However, the functions of the insulin pump are not dependent on the functionality of the remote control device, so the study suggests that insulin pump users can move safely under the power lines.
Key words: electromagnetic field / exposure / personnel / radiation / non-ionizing
© SFRP, 2020
Current usage metrics show cumulative count of Article Views (full-text article views including HTML views, PDF and ePub downloads, according to the available data) and Abstracts Views on Vision4Press platform.
Data correspond to usage on the plateform after 2015. The current usage metrics is available 48-96 hours after online publication and is updated daily on week days.
Initial download of the metrics may take a while.