Optimisation d’un protocole rochers au scanner
Optimization of a computed tomography (CT) scan protocol of the temporal bone
1
Pôle de Physique Médicale, C2i santé, Cabinet de Consultants en Ingénierie de la Santé,
10 rue Paul Langevin – Parc d’Activités St Jacques II,
54320
Maxéville, France
2
Imagerie Médicale de la Plaine de France,
7 avenue Henri Barbusse,
93150
Le Blanc Mesnil, France
3
Centre de Recherche en Automatique de Nancy (CRAN – UMR 7039), Université de Lorraine – CNRS,
54500
Vandoeuvre-les-Nancy, France
* Auteur de correspondance : hanane.miloudi.hm@gmail.com
Reçu :
18
Avril
2017
Accepté :
12
Décembre
2017
Cette étude vise à optimiser le protocole des rochers d’un scanner Siemens Somatom Sensation 40. Ainsi, plusieurs acquisitions du fantôme Catphan 600 sont réalisées pour évaluer quantitativement la qualité d’image. Pour cela, l’écart type du nombre CT et la fonction de transfert de modulation (FTM) sont mesurés à partir des images acquises. La qualité d’image est également évaluée visuellement par le radiologue, à partir d’images issues de patients. Trois critères sont considérés : la visualisation du tympan, la visualisation des osselets et la texture du bruit. L’utilisation de la table plutôt que de la têtière comme support conduit à une augmentation moyenne du bruit de 10 %. L’évaluation qualitative des images issues de patients a permis quant à elle de réduire les mAs effectifs de 28 % et donc l’IDSV de 150,8 mGy à 109,1 mGy. De plus, l’utilisation d’un filtre plus dur a conduit à une amélioration de la visualisation des plus petites structures de l’oreille interne. Ainsi, cette étude a permis d’assurer une qualité d’image satisfaisante pour le diagnostic tout en réduisant la dose délivrée au patient de 28 %.
Abstract
This study aims to optimize the CT scan protocol for temporal bone for a Somatom Sensation 40CT scanner by SIEMENS. Several acquisitions of Catphan 600 phantom are realized in order to quantitatively assess the image quality. The standard deviation of the CT Hounsfield Unit and the Modulation Transfer Function (MTF) are measured from the images acquired. The image quality is visually evaluated by the radiologist thanks to the patient images considering three criteria: visualization of the eardrum, visualization of the ossicles and noise texture. Using the cradle as the main support, in place of the headrest, a noise increment of around 10% was observed. Qualitative assessments of the patient images allowed a reduction of the effective mAs of 28% and therefore the volumetric computed tomography dose index (CTDIvol) from 150.8 mGy to 109.1 mGy. In addition, the use of a harder filter led to improved visualization of the smallest structures of the inner ear. The study results allowed to find the best compromise between the highest image quality required to realize a confident diagnostic and a considerable reduction of the dose delivered to the patient (of 28%).
Key words: computed tomography / temporal bone / patient, image quality / dose