FAZO : un programme d'analyse des spectres de thermoluminescence par la methode de déconvolution
FAZO : A thermoluminescence glow curve analysis program by the deconvolution method
Centre de Radioprotection et de Sûreté, Département de Dosimétrie des Rayonnements, Laboratoire de Dosimétrie Externe, 02, bd Frantz Fanon, B.P. 399, Alger-Gare, 16000 Alger, Algérie
FAZO est un programme informatique de déconvolution des spectres de thermoluminescence, fondé sur l'algorithme de minimisation de fonctions de Levenberg-Marquardt avec les modifications apportées par Fletcher. Elaboré et contrôlé au sein du laboratoire de dosimétrie externe du Centre de radioprotection et de s&ucric;reté (CRS, Alger, Algérie), le programme FAZO est actuellement utilisé pour la surveillance radiologique des personnels et de l'environnement à l'aide de détecteurs thermoluminescents. Dans ce travail, sont présentées les performances de FAZO validées à l'aide de spectres de thermoluminescence simulés obtenus à partir des équations cinétiques de Randall et Wilkins ainsi que par une confrontation avec les résultats d'autres programmes élaborés dans des laboratoires ayant participé à une campagne d'intercomparaison des programmes d'analyse des spectres de thermoluminescence. Il est aussi présenté des exemples de déconvolution des spectres expérimentaux du détecteur LiF:Mg,Ti (TLD 100 et TLD 700) irradié à 10 mGy et 20 Gy et du détecteur CaF2:Dy (TLD 200) exposé . 10 mGy. Enfin, les résulats obtenus par FAZO pour l'évaluation des paramètres cinétiques, l'énergie d'activation E et le facteur de fréquence s, du matériau LiF:Mg,Ti sont en bon accord avec les valeurs les plus répandues dans la littérature consacrée à ce sujet.
Abstract
FAZO : A thermoluminescence glow curve analysis program by the deconvolution method. FAZO is a computer program of thermoluminescence glow curve deconvolution based on the modified Flectcher's minimization algorithm of Levenberg- Marquardt. The program FAZO which has been elaborated and tested in the external dosimetry laboratory of the Radiation Protection and Safety Center (CRS, Algiers, Algeria) is now used for the dosimetric survey of the personnel and the gamma outdoor by means of thermoluminescent detectors. In this work, the presented performances of FAZO have been validated using simulated thermoluminescence glow curves obtained by the kinetic equation of Randall and Wilkins, and by a confrontation with the results of other programs elaborated by laboratories which have participated to a thermoluminescence glow curve analysis computer programs intercomparison. Deconvolution examples of experimental glow curves of LiF:Mg,Ti (TLD 100 and TLD 700) detector irradiated at 10 mGy and 20 Gy and CaF2:Dy (TLD 200) detector exposed to 10 mGy, are also presented in this paper. Finally, the obtained results concerning the kinetic parameters evaluation by FAZO, the activation energy E and the frequency factor s, for LiF:Mg,Ti material are found to be in good agreement with most values published on this subject.
Key words: thermoluminescence dosimetry / computer calculations
© EDP Sciences, 1998