Accès gratuit
Numéro
Radioprotection
Volume 53, Numéro 2, April-June 2018
Page(s) 139 - 144
DOI https://doi.org/10.1051/radiopro/2018012
Publié en ligne 3 mai 2018

© EDP Sciences 2018

1 Introduction

Recently, digital radiography systems are quickly replacing the conventional film–screen systems in many radiology departments around the world. Digital radiography systems have advantages including a wide dynamic range, flexibility in image display, possibility in changing image quality parameters, digital image management by using Picture Archiving and Communication System (PACS) and then reduction of costs associated with processing, managing, and storing films (Fischmann et al., 2005). However, in the digital radiography systems, large amounts of exposure can be compensated by detector-computer system and then it is relatively easy to unknowingly overexpose the patient and increase the risk of effects induced by ionizing radiation (Nahangi and Chaparian, 2015).

There is a trade-off between the radiation exposure to the patient and image quality especially in digital radiographies. Based on the principle of “as low as reasonably achievable” (ALARA), digital radiographies should provide image quality adequate to enable an accurate diagnosis with the lowest achievable radiation dose (Khong et al., 2013). Detectors with higher detection quantum efficiency can create better images with lower radiation exposure (Sun et al., 2012).

Several studies have compared images obtained with DR or computed radiography (CR) with conventional screen film radiographs (Fink et al., 2002; Fischbach et al., 2002; Bacher et al., 2003; Lu et al., 2003, Fischmann et al., 2005; Muhogora et al., 2011; Sun et al., 2012), but few studies compared different models of DR systems (Strotzer et al., 2002). Therefore, the purpose of this study was to compare image quality parameters (in terms of spatial resolution, contrast resolution, and contrast-to-noise ratio) as well as patient dose (in terms of dose area product (DAP) values) in different digital radiography systems using an appropriate phantom.

2 Material and methods

2.1 Imaging systems

In this survey, image quality and patient dose (DAP) were compared among four different digital systems. These systems are representative of the X-ray systems being used in our region. Information and technical specifications of the studied systems were summarized in Table 1. Imaging parameters were selected with exposure time of 20 ms, tube current of 100, 250, 360, 400, 560 mA, and tube voltage of 50, 60, 70, 75, 80 kVp, respectively. Three images were obtained in each setting. Source to image distance was set at 100 cm for all exposures. A total of 300 images were obtained using above parameters (5 tube current stations × 5 tube voltage stations × 3 images × 4 systems).

Table 1

Information and technical specifications of the four different digital systems installed in four hospital investigated in this study.

2.2 Pehamed FLUORAD 30 A + D phantom

The phantom used in this study was Pehamed FLUORAD 30 A + D. The picture of phantom is shown in Figure 1. Spatial resolution and contrast resolution can be measured by this phantom. The overall dimensions of the test phantom were 300 mm × 300 mm × 18.5 mm. It included a 1.5 mm thick copper plate with the details objects which was embedded between PMMA plates; The color of phantom was white, so that the collimated light field could also be seen under unfavorable lighting conditions in the X-rays room. There were 8 low contrast objects with 10 mm diameter for determinning of contrast resolution and there were a bar pattern (0.5 to 5 lp/mm) rotated 45° for determining of spatial resolution.

thumbnail Fig. 1

Image of Pehamed FLUORAD A + D.

2.3 Measurement of dose area product

The dose area product (DAP) values were measured by a DAP meter (VacuDAP OEM, VacuTec, Germany) mounted on the X-ray system and displayed on the console monitor. The displayed DAP values were recorded during image acquisition for different imaging parameters in all radiography systems. The accuracy of DAP values were checked with the calibrated DAP meter (PTW, Freiburg, Germany).

2.4 Image evaluation

Phantom images were presented to the three experienced specialists in digital radiography at the same time in random order. Each specialist analyzed image quality independently on a computer workstation. Image analysis was done in the best viewing circumstance and there was no restriction on using magnifying tools and viewing distance. The number of low contrast objects as a scale of contrast resolution, and the number of resolution lead bar pattern as spatial resolution were measured on each image. Then, the mean values and standard deviations of each parameter were calculated based on the three specialists reports.

2.5 Measurement of contrast-to-noise ratio

Contrast-to-noise ratios for all the images were measured according to the following procedure:

  • two regions of interest (ROI) should be defined in all images (original images); one of these regions was background of the image and another ROI was chosen as seen in Figure 2;

  • mean pixel value (MPV) and standard deviations (SD) of all ROI were determined with imageJ 1.44p software;

  • contrast-to-noise ratio was measured according to the following equation (Muhogora et al., 2011):

(1)

thumbnail Fig. 2

Region of interest (ROI) to measure contrast-to-noise ratio.

3 Results

The changes of the DAP values as a function of different exposure parameters (beam voltage (kVp) and tube current (mA)) for four digital radiography systems are shown in Figures 36. Considering these charts, it is clear that the DAP values increase with increasing the exposure parameters. The highest DAP values were observed at 80 kVp while the lowest DAP were observed at 50 kVp for all digital radiography systems.

Figures 79 indicate that spatial resolution, contrast resolution and contrast-to-noise ratio are different between digital systems. All three quantities increased with increasing imaging parameters in all systems. The CNR in the SEDECAL system was higher than that in other systems (p ≤ 0.05) (Fig. 7). At the stable DAP (100 mGy.cm2), the highest mean of CNR (29.70 ± 0.85) was related to SEDECAL system (p ≤ 0.05). Figure 8 represents that the spatial resolution at fixed DAP of 100 mGy.cm2 for the SEDECAL and Applem- R302/A DR systems were 3.10 ± 0.38 lp/mm and 3.12 ± 0.11 lp/mm, respectively, which were significantly higher than that in other systems (p ≤ 0.05). As it is shown in Figure 9, SEDECAL system has also significantly higher CR at stable DAP (100 mGy.cm2); 6.38 ± 0.797 and the Applem- R302/A DR system with CR of 4.07 ± 0.42 is located on the next rank (p ≤ 0.05). Figures 79 indicate that to obtain images with a fixed image quality (CR, SR and CNR), different DAP values are needed in different digital radiography systems. Also, it is inferred that the SEDECAL system needs the lowest DAP at all image quality parameters (p < 0.05). At fixed CR of 4 and CNR of 25, DAP value obtained in SEDECAL system was lower than others (p ≤ 0.05) (Fig. 9) and at fixed SR of 3.36 lp/mm, DAP values in SEDECAL and Applem- R302/A were lower than two other systems (p ≤ 0.05) (Fig. 8).

thumbnail Fig. 3

The changes of the DAP values as a function of different exposure parameters (beam voltage (kVp) and tube current (mA)) for in Applem- R302/A Digital radiology.

thumbnail Fig. 4

The changes of the DAP values as a function of different exposure parameters (beam voltage (kVp) and tube current (mA)) for in MEHRAN TEB (B) Digital radiology.

thumbnail Fig. 5

The changes of the DAP values as a function of different exposure parameters (beam voltage (kVp) and tube current (mA)) for in MEHRAN TEB (A) Digital radiology.

thumbnail Fig. 6

The changes of the DAP values as a function of different exposure parameters (beam voltage (kVp) and tube current (mA)) for in SEDECAL Digital radiology.

thumbnail Fig. 7

The relationship between CNR and DAP in different digital radiology systems.

thumbnail Fig. 8

The relationship between SR and DAP in different digital radiology systems.

thumbnail Fig. 9

The relationship between CR and DAP in different digital radiology systems.

4 Discussion

The basic goal of optimization in X-ray examinations is to minimize the patient dose while still providing image quality adequate for correct diagnosis. The performance of each imaging equipment can infuence both image quality and the patient dose. The simplest technique for assessing the performance of each equipment is to use a phantom under ideal conditions, away from problems related to clinic. Thus, in this survey, image quality was evaluated using the phantom and, simultaneously, radiation exposure was measured. There are two important findings in this survey: Firstly, different digital radiography systems have different performance. So, imaging parameters for one system should not be used for others and imaging parameters should be optimized individually for each system. Secondly, a system with the best performance that produces better image quality at lower doses was identified.

In this survey, the diagnostic performance of four different digital radiography systems has been assessed in terms of image quality and DAP using the Pehamed phantom. While, CR and SR assessments were based on observation of three experienced specialists, CNR evaluations were based on calculations performed by the imageJ software. Image quality assessments in other studies were also based on either observations of specialists (Fink et al., 2002; Bliznakova et al., 2003; Fischmann et al., 2005) or CNR calculations (Hess and Neitzel, 2011, 2012; Mori et al., 2013) or both methods (Saarakkala et al., 2009; Muhogora et al., 2011).

In this study, it was demonstrated that different digital systems have different performance. The reason for these results is that CR and CNR are almost similar criterias for image quality assessment and influenced by same pararmeters such as contrast characteristics of the receptor and display system, but SR is influenced by dexel size of detector and by factors unrelated to the image receptor such as focal spot size, pixel size of displying monitor. In the survey conducted by Sun et al. (2012), similar conclusion have been reported. Of course, they assessed the performance of three computed radiography and three direct radiography systems in terms of their image noise and entrance skin dose.

This study suffers from some limitations. The major limitation of this study was that a comparison of the current and the previous surveys was difficult because there were few studies that evaluated the performance of different digital radiography systems. Another limitation was that, there were few models of digital radiography in this study. Further work is required to evaluate other models of digital radiography systems to complete these results.

5 Conclusion

This study aimed to apply a simple method based on phantom measurements to find the best digital system model among the four DR systems. Overall performance of SEDECAL digital radiography system is superior to that of other systems. SEDECAL system has higher CNR, CR and SR at stable DAP and and vice versa lower DAP at stable image quality parameters. The results of this investigation can be taken into consideration in the selection and purchasing of new systems in order to preserve patients as well as radiographers from unnecessary dose consistent with the radiation protection principle ALARA.

Acknowledgments

The authors would like to thank the assistance of the radiographers from the following hospitals at Yazd for our study: Shahid Sadoghi Hospital, Shahid Rahnamoon (Farokhi) Hospital, Fateme Al-Zahra Hospital, and Afshar Hospital.

References

  • Bacher K, Smeets P, Bonnarens K, De Hauwere A, Verstraete K, Thierens H. 2003. Dose reduction in patients undergoing chest imaging: digital amorphous silicon flat-panel detector radiography versus conventional film-screen radiography and phosphor-based computed radiography. Am. J. Roentgenol. 181(4): 923–929. [CrossRef] [Google Scholar]
  • Bliznakova K, Bliznakov Z, Bravou V, Kolitsi Z, Pallikarakis N. 2003. A three-dimensional breast software phantom for mammography simulation. Phys. Med. Biol. 48(22): 3699. [CrossRef] [PubMed] [Google Scholar]
  • Fink C, Hallscheidt PJ, Noeldge G, Kampschulte A, Radeleff B, Hosch WP, Kauffmann GW, Hansmann J. 2002. Clinical comparative study with a large-area amorphous silicon flat-panel detector: image quality and visibility of anatomic structures on chest radiography. Am. J. Roentgenol. 178(2): 481–486. [CrossRef] [Google Scholar]
  • Fischbach F, Ricke J, Freund T, Werk M, Spors B, Baumann C, Pech M, Felix R. 2002. Flat panel digital radiography compared with storage phosphor computed radiography: assessment of dose versus image quality in phantom studies. Invest. Radiol. 37(11): 609–614. [CrossRef] [PubMed] [Google Scholar]
  • Fischmann A, Siegmann K, Wersebe A, Claussen C, Muller-Schimpfle M. 2005. Comparison of full-field digital mammography and film-screen mammography: image quality and lesion detection. Br. J. Radiol. 78(928): 312–315. [CrossRef] [PubMed] [Google Scholar]
  • Hess R, Neitzel U. 2011. Optimizing image quality and dose in digital radiography of pediatric extremities [Internet]. Philips Healthcare. Available from http://www.healthcare.philips.com/main/about/events/rsna/pdfs/DR_White_paper_Optimizing_image_quality_and_dose_in_digital_radiography_of_pediatric_extremities.pdf. [Google Scholar]
  • Hess R, Neitzel U. 2012. Optimizing image quality and dose for digital radiography of distal pediatric extremities using the contrast-to-noise ratio. Rofo 184(7): 643–649. [Google Scholar]
  • Khong P, Ringertz H, Donoghue V, Frush D, Rehani M, Appelgate K, Sanchez R. 2013. ICRP publication 121: radiological protection in paediatric diagnostic and interventional radiology. Ann. ICRP 42(2): 1–63. [CrossRef] [Google Scholar]
  • Lu Z, Nickoloff E, So J, Dutta A. 2003. Comparison of computed radiography and film/screen combination using a contrast-detail phantom. J. Appl. Clin. Med. Phys. 4(1): 91–98. [CrossRef] [PubMed] [Google Scholar]
  • Mori M, Imai K, Ikeda M, Iida Y, Ito F, Yoneda K, Enchi Y. 2013. Method of measuring contrast-to-noise ratio (CNR) in nonuniform image area in digital radiography. Electron. Commun. Jpn. 96(7): 32–41. [CrossRef] [Google Scholar]
  • Muhogora W, Padovani R, Msaki P. 2011. Initial quality performance results using a phantom to simulate chest computed radiography. J. Med. Phys./Assoc. Med. Phys. India 36(1): 22. [Google Scholar]
  • Nahangi H, Chaparian A. 2015. Assessment of radiation risk to pediatric patients undergoing conventional X-ray examinations. Radioprotection 50(1): 19–25. [CrossRef] [EDP Sciences] [Google Scholar]
  • Saarakkala S, Nironen K, Hermunen H, Aarnio J, Heikkinen J. 2009. Comprehensive optimization process of paranasal sinus radiography. Acta Radiol. 50(3): 327–333. [CrossRef] [PubMed] [Google Scholar]
  • Strotzer M, Völk M, Fründ Rd, Hamer O, Zorger N, Feuerbach S. 2002. Routine chest radiography using a flat-panel detector: image quality at standard detector dose and 33% dose reduction. Am. J. Roentgenol. 178(1): 169–171. [CrossRef] [Google Scholar]
  • Sun Z, Lin C, Tyan Y, Ng K-H. 2012. Optimization of chest radiographic imaging parameters: a comparison of image quality and entrance skin dose for digital chest radiography systems. Clin. Imaging 36(4): 279–286. [CrossRef] [PubMed] [Google Scholar]

Cite this article as: Kaheni H, Chaparian A, Nafisi-Moghadam R, Hamzian N. 2018. Assessment of image quality and radiation dose in some models of digital radiography systems – A Pehamed FLUORAD A + D phantom study. Radioprotection 53(2): 139–144

All Tables

Table 1

Information and technical specifications of the four different digital systems installed in four hospital investigated in this study.

All Figures

thumbnail Fig. 1

Image of Pehamed FLUORAD A + D.

In the text
thumbnail Fig. 2

Region of interest (ROI) to measure contrast-to-noise ratio.

In the text
thumbnail Fig. 3

The changes of the DAP values as a function of different exposure parameters (beam voltage (kVp) and tube current (mA)) for in Applem- R302/A Digital radiology.

In the text
thumbnail Fig. 4

The changes of the DAP values as a function of different exposure parameters (beam voltage (kVp) and tube current (mA)) for in MEHRAN TEB (B) Digital radiology.

In the text
thumbnail Fig. 5

The changes of the DAP values as a function of different exposure parameters (beam voltage (kVp) and tube current (mA)) for in MEHRAN TEB (A) Digital radiology.

In the text
thumbnail Fig. 6

The changes of the DAP values as a function of different exposure parameters (beam voltage (kVp) and tube current (mA)) for in SEDECAL Digital radiology.

In the text
thumbnail Fig. 7

The relationship between CNR and DAP in different digital radiology systems.

In the text
thumbnail Fig. 8

The relationship between SR and DAP in different digital radiology systems.

In the text
thumbnail Fig. 9

The relationship between CR and DAP in different digital radiology systems.

In the text

Les statistiques affichées correspondent au cumul d'une part des vues des résumés de l'article et d'autre part des vues et téléchargements de l'article plein-texte (PDF, Full-HTML, ePub... selon les formats disponibles) sur la platefome Vision4Press.

Les statistiques sont disponibles avec un délai de 48 à 96 heures et sont mises à jour quotidiennement en semaine.

Le chargement des statistiques peut être long.