Comparison of water-equivalent and effective diameter-based size-specific dose estimates in computed tomography pulmonary angiography: implications for radiation dose optimization
1
Hassan 1st University Settat, Laboratory of Health Sciences and Technologies, High Institute of Health Sciences, Settat, Morocco
2
Department of Radiology Public Hospital, Mediouna, Morocco
* Corresponding author: a.khallouqi@uhp.ac.ma
Received:
26
September
2024
Accepted:
16
December
2024
This retrospective study investigates dose estimation in computed tomography pulmonary angiography (CTPA) by analyzing 200 examinations from 96 male and 104 female patients. Two size-specific dose estimation (SSDE) methods were evaluated: one based on effective diameter (SSDEDeff) and the other on water-equivalent diameter (SSDEDw). The effective diameter (Deff) was manually measured using the lateral (LAT) and anteroposterior (AP) dimensions, while the water-equivalent diameter (Dw) was calculated from the region of interest (ROI) using mean Hounsfield Unit (HU) values and cross-sectional area. SSDE calculations were compared to the conventional CT dose metric, CTDIvol, to assess its underestimation relative to SSDEDw. The findings revealed a dose underestimation of up to 18% for CTDIvol compared to SSDEDw, particularly for patients with larger body habitus. SSDEDeff consistently overestimated the dose by 7.73% compared to SSDEDw, as Deff relied only on external dimensions while Dw considered tissue attenuation for a more individualized dose assessment. High correlations were observed between lateral diameter (dLAT) and both Deff (R2 = 0.9175) and Dw (R2 = 0.7578). However, the systematic overestimation by SSDEDeff emphasizes the importance of clearly specifying the metric used for SSDE, as differences can influence reported doses by 5–10%, affecting clinical dose monitoring and adherence to diagnostic reference levels. This study highlights the limitations of CTDIvol and SSDEDeff compared to SSDEDw, especially in regions with heterogeneous tissue. The results support the use of individualized radiation dose assessments to improve radiation safety, particularly in CTPA, which has seen increased application during the COVID-19 pandemic.
Key words: CTPA / radiation dose / size-specific dose estimates / water-equivalent diameter / COVID-19 / Morocco
© A. Khallouqi et al., Published by EDP Sciences 2025
This is an Open Access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
Les statistiques affichées correspondent au cumul d'une part des vues des résumés de l'article et d'autre part des vues et téléchargements de l'article plein-texte (PDF, Full-HTML, ePub... selon les formats disponibles) sur la platefome Vision4Press.
Les statistiques sont disponibles avec un délai de 48 à 96 heures et sont mises à jour quotidiennement en semaine.
Le chargement des statistiques peut être long.