Dose measurement study of the chest during a CT scan using a fabricated phantom
Department of Radiological Sciences, Applied Medical Sciences College, Najran University, Najran, Saudi Arabia
* Corresponding author: ffalqahtani@nu.edu.sa
Received:
12
September
2022
Accepted:
13
December
2022
The purpose of this study was to evaluate radiation doses during computer tomography (CT) scans of the whole body and organs using a fabricated phantom. The Weighted Computer Tomography Dose Index (CTDIw) was evaluated and measured using a fabricated phantom, and five CT scanners were used for chest examinations from four hospitals in Najran, Saudi Arabia. The heart, lungs, esophagus, and spinal cord doses were measured with clinical whole body CT acquisition protocol using Thermoluminescent Dosimeters (TLDs). The TLDs used for measurements were selected for an accuracy of 5% and calibrated for the X-ray radiation field. The CTDIw and organ doses reported in this study were compared with results of ImPACT software for calculating organ doses. The results show good agreement with the calculations and the literature. The measured specific organ doses ranged between 1.33 mGy (lung marrow) and 1.65 mGy (esophagus). The model is easy to produce and useful for evaluating the organ doses and CTDIw.
Key words: fabricated phantom / computed tomography / radiation dose / TLD
© SFRP, 2023
Les statistiques affichées correspondent au cumul d'une part des vues des résumés de l'article et d'autre part des vues et téléchargements de l'article plein-texte (PDF, Full-HTML, ePub... selon les formats disponibles) sur la platefome Vision4Press.
Les statistiques sont disponibles avec un délai de 48 à 96 heures et sont mises à jour quotidiennement en semaine.
Le chargement des statistiques peut être long.