Open Access

This article has an erratum: [https://doi.org/10.1051/radiopro/2025012]


Table 1

Data obtained in this study, including mean, minimum, and maximum, as well as percentiles for CTDIvol (mGy) and DLP (mGy.cm) numbers found.

Exposure parameters CT exams

  Head CAP Chest Abdomen
Voltage (kV) [120–120] [120–120] [120–120] [120–120]
Intensity (mAs) [270–270] [45–180] [48–244] [17–71]
Thickness (mm) [5–10] [5–10] [5–10] [5–10]
Pitch [0.55–1.5] [0.55–1.5] [0.55–1.5] [0.55–1.5]
Mean CTDIvol (mGy) 38 7 9 7
Min CTDIvol (mGy) 36 3 3 5
Max CTDIvol (mGy) 44 12 17 12
50th CTDIvol (mGy) 34 6 7 7
75th CTDIvol (mGy) 40 8 11 9
Mean DLP (mGy.cm) 1422 710 598 454
Min DLP (mGy.cm) 1281 354 333 121
Max DLP (mGy.cm) 1724 1222 874 1021
50th DLP (mGy.cm) 1390 695 612 704
75th DLP (mGy.cm) 1445 825 700 788

Current usage metrics show cumulative count of Article Views (full-text article views including HTML views, PDF and ePub downloads, according to the available data) and Abstracts Views on Vision4Press platform.

Data correspond to usage on the plateform after 2015. The current usage metrics is available 48-96 hours after online publication and is updated daily on week days.

Initial download of the metrics may take a while.