Free Access

Table 3

Summary of entrance skin dose of three cardiologists in six different points and radiation exposure parameters due to angiography procedures.

  Median ± SD (range)
Variable Physician 1 Physician 2 Physician 3
Number of views 6 ± 2.4 (3–11) 7.5 ± 4.6 (3–17) 7 ± 0.8 (6–8)
Fluoroscopy time (min) 1.1 ± 2 (0.8–7.2) 2.5 ± 4.6 (0.5–17.2) 0.8 ± 0.7 (0.5–2.3)
DAP (mGy.m2) 1.4 ± 1.5 (0.7–5.8) 2.4 ± 1.8 (0.7–7.1) 1 ± 0.6 (0.8–2.5)
IRP Dose (mGy) 212 ± 277 (91.9–1004) 351 ± 269 (127–1104) 150 ± 97 (122–376)
Thyroid dose (μGy) 0.0 (0–0.84) 0.88 (0–2.93) 0.0 (0–1.25)
Right chest dose (μGy) 0.0 (0–3.24) 0.39 (0–1.90) 0.0 (0–1.26)
Left chest dose (μGy) 0.02 (0–1.24) 1.53 (0–5.74) 0.02 (0–1.58)
Right wrist dose (μGy) 2.23 (0.31–9.42) 5.55 (0.83–10.81) 1.93 (0–3.38)
Left wrist dose (μGy) 13.80 (0.51–49.15) 13.54 (8.71–24.73) 9.50 (3.44–25.08)
Left leg dose (μGy) 29.25 (0.70–128.61) 46.58 (5.81–155.14) 19.11 (2.03–44.23)

All data presented as median, SD and range.

Current usage metrics show cumulative count of Article Views (full-text article views including HTML views, PDF and ePub downloads, according to the available data) and Abstracts Views on Vision4Press platform.

Data correspond to usage on the plateform after 2015. The current usage metrics is available 48-96 hours after online publication and is updated daily on week days.

Initial download of the metrics may take a while.