Free Access

Table 1

Comparison between calculated (present work), experimental (Byun et al., 2004) and simulated (Byun et al., 2004) total efficiencies for an isotropic radiating axial point sources placed at the center of a 4πNaI(Tl) (30.6 × 30.6 × 40.6 cm3) detector with a central rectangle hole (10.2 × 10.2 × 40.6 cm3).

Radioisotope Eγ  (MeV) Exp. (Byun et al., 2004) M.C.
(Byun et al., 2004)
Present work Δ1% Δ2%
22Na 0.511 0.912 ± 0.002 0.911 ± 0.001 0.901 −1.221 −1.110
137Cs 0.662 0.947 ± 0.028 0.896 ± 0.001 0.890 −6.404 −0.674
60Co 1.173 0.861 ± 0.003 0.845 ± 0.001 0.867 0.692 2.538
22Na 1.275 0.857 ± 0.003 0.837 ± 0.001 0.860 0.349 2.674
60Co 1.332 0.854 ± 0.004 0.832 ± 0.001 0.858 0.466 3.030

Current usage metrics show cumulative count of Article Views (full-text article views including HTML views, PDF and ePub downloads, according to the available data) and Abstracts Views on Vision4Press platform.

Data correspond to usage on the plateform after 2015. The current usage metrics is available 48-96 hours after online publication and is updated daily on week days.

Initial download of the metrics may take a while.